A field service engineer (fse) was at the customer site to address the reported issue.The fse confirmed the reported error on the error log and reproduced the issue by attempting to run a control sample.The fse resolved the issue by adjusting clog parameters back into specification as it was found to be off and adjusted alignment to the diluent well.Instrument validation was performed via quality control (qc).Qc results passed within the manufacturer's published ranges.The aia-900 instrument returned to normal operation.No further field action required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2019 to aware date (b)(6) 2020.No other similar complaints were identified during the search period.The aia-900 operator's manual states the following: [2076] clogging detected at specimen cause: the negative pressure generated by suction of a (rack id, rack, position, specimen id) specimen exceeded the standard value.There is a possibility that the sampling nozzle was clogged, preventing the specified quantity of specimen suction from taking place.An sc flag will be attached to the measurement result.Action: if there is no problem with the specimen, contact the tosoh local representatives.The probable cause of the issue is attributed to clog parameters were out of specification and misalignment of the diluent well.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|
A customer reported receiving error 2076 clogging detected at specimen, in addition to clogging detected at sds and pretreatment, on the aia-900 analyzer.The error occurred when the customer attempted to run vitamin b12 and vitamin d.As part of troubleshooting, the customer cleaned the sample probe then ran free thyroxine (ft4).The error occurred and the instrument was not operational.A field service engineer (fse) was notified.A field service engineer (fse) was dispatched to address the reported issue, which resulted in the delay of reporting intact parathyroid hormone (ipth) patient results.There was no report of patient intervention or adverse health consequences due to delay in reporting.
|