Model Number N/A
Device Problem Material Frayed (1262)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that there are exposed wires during testing.There was no harm or delay.The date is unknown.There was no adverse event reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the there were exposed wires.The plug harness assembly was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
There is no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key10355182
MDR Text Key201556809
Report Number0001526350-2020-00644
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63592362
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A