• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX COLONIC; STENT, COLONIC, METALIC, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX COLONIC; STENT, COLONIC, METALIC, EXPANDABLE Back to Search Results
Model Number M00565050
Device Problem Activation Failure (3270)
Patient Problem No Code Available (3191)
Event Date 06/21/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).The complainant indicated that the suspect device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation on july 13, 2020 that a wallflex colonic stent was implanted to treat a colonic stricture in the descending colon during a colonic stent procedure performed on (b)(6) 2020.Reportedly, the patient's anatomy was not tortuous and was not dilated prior to stent placement.According to the complainant, on (b)(6) 2020 during a routine post stent placement examination, the physician noticed that the stent did not fully expand.Reportedly, the physician was comfortable leaving the stent implanted; however, the physican plans to expand the stent with a balloon dilator.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX COLONIC
Type of Device
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key10355618
MDR Text Key207509983
Report Number3005099803-2020-03094
Device Sequence Number1
Product Code MQR
UDI-Device Identifier08714729456520
UDI-Public08714729456520
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K061877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2021
Device Model NumberM00565050
Device Catalogue Number6505
Device Lot Number0023192151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/13/2020
Initial Date FDA Received08/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-