• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BHR RESURFACING FEMORAL HEAD 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BHR RESURFACING FEMORAL HEAD 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74123152
Device Problems Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Injury (2348); Metal Related Pathology (4530)
Event Date 02/12/2020
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed on the patients right hip due to pain, disability, and mental suffering.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review and device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.Smith and nephew has not received the device/adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
H3, h6: it was reported that a right hip revision surgery was performed due to pain, stiffness, swelling and metallosis.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the cup and the head.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the cup and the head, and this failure will continue to be monitored.As no device batch numbers were provided for investigation, manufacturing record review could not be performed.If more information is received, this investigation will be reopened.The review of the most recent ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.The available medical documents were reviewed.Although, the reported pain, stiffness, swelling, elevated metal ion levels, as well as the intraoperative findings of dark fluid and tissue and detached, necrotic abductors may be consistent with findings associated with metallosis; however, the root cause cannot be confirmed and it cannot be concluded that the reported reactions were associated with a mal-performance of the implant.The patient impact beyond the reported events cannot be determined.However, it was reported that the patient still experiences weakness and instability.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.H11: corrected data b1: adverse event and product problem e1: initial reporter name and address g- contact office g2- report source.
 
Event Description
It was reported that a right revision surgery was performed on (b)(6) 2020 due to pain, stiffness, swelling and metallosis.During the surgery, all the components were explanted and replaced with a tha system from a competitor manufacturer.The primary right hip resurfacing surgery was performed on (b)(6) 2010.The patient still has weakness and instability.
 
Manufacturer Narrative
H10.Additional information was added in h6 (health effect - impact code).H11.Corrected information in h6 (medical device problem code & health effect - clinical code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR RESURFACING FEMORAL HEAD 52MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
spa park
austin, TX 
5123913905
MDR Report Key10355917
MDR Text Key201352195
Report Number3005975929-2020-00276
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010552433
UDI-Public03596010552433
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number74123152
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? No
Date Manufacturer Received09/01/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexMale
Patient Weight109 KG
-
-