Model Number N/A |
Device Problems
Failure to Cut (2587); Intermittent Loss of Power (4016)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/17/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/ final report will be submitted.
|
|
Event Description
|
It was reported that the dermatome doesn't cut possibly due to motor failure.The inspected dermatome motor didn't work properly during setup (before cutting).There was no patient harm, delay unknown, and no alternate contact.No adverse event was reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor was replaced and resolved the reported issue.Review of the previous repair record identified the motor was replaced, which is related to the reported event.The device was tested and found to be functioning to specification prior to release to the customer.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|