MAUDE Adverse Event Report: ARTHROCARE CORP. DISPOSABLE FIRSTPASS SUTURE PASSER, SELF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DISPOSABLE FIRSTPASS SUTURE PASSER, SELF Back to Search Results
Catalog Number 22-4038
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 07/09/2020
Event Type  malfunction  
Event Description
It was reported that when using the device during a shoulder surgery, the retractable needle did not retract when it was located at the shoulder cuff.The surgeon was able to remove the medical device carefully without severing the shoulder cuff.The procedure was completed with the same device.It is unknown if there was a delay in the case.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event Description
It was reported that when using the device during a shoulder surgery, the retractable needle did not retract when it was located at the shoulder cuff.The surgeon was able to remove the medical device carefully without severing the shoulder cuff across the surgical approach.However, besides the needle perforation, there was damage to subcutaneous and muscle non-target tissues.The procedure was completed with the same device and no significant delay was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10.Additional information in b5.
 
Manufacturer Narrative
The reported device, used in treatment, was returned for evaluation.There was a relationship found between the reported incident and the returned device.A review of the device history records for the reported lot number showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found: -as with any surgical instrument, care should be taken to ensure that excessive force is not placed on the device.- do not use this device as a lever for manipulating hard tissue or bone.Excessive force should not be applied to the device when manipulating soft tissue, bone, or hard objects.Misuse of this device may result in bent or broken distal tip or jaws.- when passing suture multiple times always check the tip of the device for damage or debris between passes.Clear any visible debris between passes.Discontinue use if the device becomes damaged between passes.-tissue thickness may affect suture placement including stitch depth and needle entry point.-ensure adequate visibility when using the device.Should the needle tip become lodged (e.G., in bone), the needle should be disengaged by retracting it into the suture passer.Twisting or bending of the needle to dislodge the needle tip may result in breakage of the needle and needle parts may not be retrievable.Visual inspection shows no manufacturing abnormalities on the device.The instrument was returned with a closed top bracket which could not be opened; the needle is bent.The first step of the trigger is stuck; releasing is not possible; the device was dissected and a jam was indicated; a bent pin was found ; further tests were not possible due to the damage.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness.(3) damage or debris in the device tip between passes.No containment or corrective actions are recommended at this time.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE FIRSTPASS SUTURE PASSER, SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10356926
MDR Text Key201404019
Report Number3006524618-2020-00568
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885556724545
UDI-Public885556724545
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup,Followup
Report Date 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/08/2022
Device Catalogue Number22-4038
Device Lot Number2042054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Date Manufacturer Received09/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight85
-
-