ST. JUDE MEDICAL PUERTO RICO, B.V. (CD) SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGFN-756
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/07/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020 a patient underwent an aortic valve replacement procedure (device: regent heart valve, model: 19agfn-756, sn: (b)(4).During the procedure, the physician found the valve blade switch to be "stuck." the physician replaced the device with a new valve product to complete the procedure.The patient is recovering well now.
 
Manufacturer Narrative
The reported event of restricted leaflet movement could not be confirmed.Morphological and hydrodynamic examination indicated the valve met abbott specifications.Hydrodynamic testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, B.V. (CD)
lot 21
caguas west industrial park
caguas, puerto rico 00726 -099
MDR Report Key10359571
MDR Text Key202383668
Report Number3003681312-2020-00009
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 08/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number7090857
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received08/13/2020
Supplement Dates FDA Received08/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-