• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS; OQY INTRAUTERINE BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS; OQY INTRAUTERINE BALLOON Back to Search Results
Model Number G24237
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It is reported during an unspecified procedure using a cook bakri postpartum balloon with rapid instillation components, the balloon was deflating (losing saline) with a visible decrease in balloon volume.It is not known how the procedure was completed after this difficulty.The complaint device is not available for physical evaluation.It was discarded by the user facility.No adverse effects to the patient have been reported as a result of this alleged product malfunction.Additional details regarding the patient and event have been requested.At this time, no additional information has been provided.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: section c.Event summary: it is reported during an unspecified procedure using a cook bakri postpartum balloon with rapid instillation components, the balloon was deflating (losing saline) with a visible decrease in balloon volume.It is not known how the procedure was completed after this difficulty.No adverse effects to the patient have been reported as a result of this alleged product malfunction.Investigation - evaluation.Reviews of the complaint history, device history record, instructions for use, and quality control procedures of the device were conducted during the investigation.No device was returned for investigation.Accordingly, no physical examinations were performed.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.There were no identified gaps in the device quality controls procedures.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is packaged with instructions for use which state, "upon removal from the package, inspect the product to ensure no damage has occurred." based on the available information, cook has concluded that a definitive cause for this event could not be determined.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS
Type of Device
OQY INTRAUTERINE BALLOON
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10361318
MDR Text Key201599524
Report Number1820334-2020-01416
Device Sequence Number1
Product Code OQY
UDI-Device Identifier10827002242378
UDI-Public(01)10827002242378(17)221017(10)10089113
Combination Product (y/n)N
PMA/PMN Number
K170622
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/17/2022
Device Model NumberG24237
Device Catalogue NumberJ-SOSR-100500
Device Lot Number10089113
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/29/2020
Initial Date FDA Received08/04/2020
Supplement Dates Manufacturer Received09/15/2020
Supplement Dates FDA Received09/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-