Model Number 0165SI14
Device Problems Inaccurate Flow Rate (1249); Obstruction of Flow (2423)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 07/10/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that no urine flow observed through the catheter on the day of use.
 
Manufacturer Narrative
The reported event was unconfirmed since the problem could not be reproduced.Evaluation found no blockage in the drainage lumen.Water was introduced through the drainage funnel and came out of the drainage eye without difficulty.The flow rate test has been performed on the returned sample and passed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[precautions] 1.Precautions for use (exercise caution when using the device in the following patients) (1) exercise caution when using the device in patients with high urinary calcium levels as encrustation on the balloon surface, catheter occlusion or damage may occur." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that no urine flow observed through the catheter on the day of use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10361358
MDR Text Key202687969
Report Number1018233-2020-04945
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016745
UDI-Public(01)00801741016745
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2023
Device Model Number0165SI14
Device Catalogue Number0165SI14
Device Lot Number9SX019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2020
Date Manufacturer Received09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-