RESMED LTD ASTRAL 150 - AMER Back to Search Results
Model Number 27003
Device Problem Failure to Read Input Signal (1581)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/06/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to resmed for an evaluation.A resmed customer representative assisted the customer with instructions to perform a hard reset of the device.The hard reset was able to resolve the issue.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to a software issue.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.(b)(4).
 
Event Description
It was reported to resmed that an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU   2153
MDR Report Key10363355
MDR Text Key202611474
Report Number3004604967-2020-00788
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Device Lot Number1122804
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date07/06/2020
Date Manufacturer Received07/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-