This event was observed on july 2, 2020 when the tech opened the scope and noticed the crack in the lens.The scope was one that had previously been repaired.There was no delay in the intended procedure and the intended procedure was completed.There was no patient harm or injuries.The same device was not used to complete the procedure.No other devices were replaced during the procedure.The user facility also informed olympus that there was no visual deformation such as kinks or buckles on the bending section of the scope.No anomalies were observed during inspection.A leak test was not performed as this is a rigid scope.The device is reprocessed per an autoclaved temp exposure time, 4 minutes.The scope is stored with the other instruments in the set.No metal was protruding, and the bending section was not broken.A device history record (dhr) did not find any defects or nonconformities.All records indicate that the device was manufactured in accordance with all applicable procedures.The probable cause for the reported event is excessive force during handling.
|