• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY UNI TIBIAL BASE RM/LL SZ 3; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY UNI TIBIAL BASE RM/LL SZ 3; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71422433
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060)
Event Date 07/02/2020
Event Type  Injury  
Event Description
It was reported that patient had smith+nephew components since 2018.A revision surgery was performed on (b)(6) 2020 due to arthrosis.The parts were ex-planted due to painful prosthesis and aseptic mobilization.
 
Manufacturer Narrative
It was reported that patient had a revision surgery was performed, 2 years post-op, due to pain, arthrosis and aseptic mobilization.The associated device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of manufacturing records did not reveal any deviation from the standard manufacturing processes.A review of the complaint history on the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Based on this investigation, the need for corrective action is not indicated.Some potential causes could include but are not limited to traumatic injury or patient conditions.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY UNI TIBIAL BASE RM/LL SZ 3
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10364530
MDR Text Key201605069
Report Number1020279-2020-03694
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
PMA/PMN Number
K102069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number71422433
Device Lot Number18DM22431
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/10/2020
Initial Date FDA Received08/04/2020
Supplement Dates Manufacturer Received07/10/2020
Supplement Dates FDA Received01/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-