Model Number 3100 A
Device Problem Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Date 07/07/2020
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).A vyaire field service representative (fsr) went onsite and evaluated the ventilator.Fsr was able to duplicate the issue with customer's circuit.Replaced circuit with test circuit of fsr and passed both pt cal and performance check.Customer's circuit either leaks or have old caps on.Advised customer to replace circuit with new style caps.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the 3100a pr3 is maxed and map is fluctuating.The customer confirmed that there was no patient harm associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, il 
3273284
MDR Report Key10365460
MDR Text Key201704377
Report Number2021710-2020-12280
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20060320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/07/2020
Initial Date FDA Received08/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-