BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 1X40G; BONE CEMENT, ANTIBIOTIC Back to Search Results
Catalog Number 3003940001
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/20/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).The device will not be returned to the manufacturer.Therefore it will not be analyzed.According to available data, the most probable root cause is due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was found that the inner sterile packaging was opened before the operation.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
(b)(4) this follow-up report is being submitted to relay additional information.The device was not returned to the manufacturer.Therefore, it could not be analyzed.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.1 complaint has been recorded over the batch a817ce2506.According to available data, the most probable root cause is due to packaging issue (sealing process).Corrective action has been initiated to address reported issue.A customer letter will be send regarding the reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It has been reported that the inner sterile packaging was found opened before the operation.No adverse events have been reported as a result of the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFOBACIN BONE CEMENT R 1X40G
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
MDR Report Key10366173
MDR Text Key201734516
Report Number3006946279-2020-00142
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
PMA/PMN Number
K150850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Catalogue Number3003940001
Device Lot NumberA817CE2506
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/25/2020
Initial Date FDA Received08/05/2020
Supplement Dates Manufacturer Received10/07/2020
Supplement Dates FDA Received10/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-