MAUDE Adverse Event Report: CONSIGNED 115V DOMESTIC HOTLINE; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONSIGNED 115V DOMESTIC HOTLINE; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number CON-HL-90
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
Information was received indicating that a smiths medical pump's lcd numbers would jump up and down.There were no reported adverse events.
 
Manufacturer Narrative
Returned device was received cracked enclosure and tank cover.Outdated pcb, power switch, and front cover.Worn line cord.During the evaluation of the device the customer reported condition was confirmed.Problem source was traced to user interface.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSIGNED 115V DOMESTIC HOTLINE
Type of Device
WARMER, THERMAL, INFUSION FLUID
MDR Report Key10367812
MDR Text Key201702792
Report Number3012307300-2020-07824
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2020
Initial Date FDA Received08/05/2020
Supplement Dates Manufacturer Received09/04/2020
Supplement Dates FDA Received09/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-