• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII MIS DCF DISTAL CUT BLK; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CMNTD,METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII MIS DCF DISTAL CUT BLK; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CMNTD,METAL/POLYMER Back to Search Results
Model Number 71441147
Device Problem Failure to Disconnect (2541)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2020
Event Type  malfunction  
Event Description
It was reported that the trocar pin got jammed in the alignment hole.It is unknown if there was a delay.No patient injuries reported.An s&n backup was not required as the procedure was finished with the same device.
 
Manufacturer Narrative
The devices, uses in treatment, were returned for evaluation.A visual inspection confirmed the pin is stuck in the cutting block.The devices show significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF DISTAL CUT BLK
Type of Device
PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CMNTD,METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10367879
MDR Text Key201706903
Report Number1020279-2020-03717
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010497185
UDI-Public03596010497185
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441147
Device Catalogue Number71441147
Device Lot Number11CM14995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Initial Date Manufacturer Received 07/13/2020
Initial Date FDA Received08/05/2020
Supplement Dates Manufacturer Received09/08/2020
Supplement Dates FDA Received09/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-