A 13-month complaint history review for similar complaints was performed for lh ii lot number jx13944 from 09jun2019 to aware date (b)(6) 2020.No other similar complaints were identified during the search period.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 09jun2019 to aware date (b)(6) 2020.No other similar complaints were identified during the search period.The st aia-pack lh ii analyte application manual states the following: evaluation of results.Quality control.In order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.This issue is under further investigation.
|