One 7f, quadripolar, large curl, xls livewire tc ablation catheter was received for evaluation.The device was returned due to a signal issue on electrode 3.Visual inspection revealed a black foreign material completely covering electrode 3 and a signal was unable to be obtained from the electrode.Electrical testing revealed a short circuit between the tip electrode and electrode 3.Dissection revealed that the flat wire insulation was torn and the insulation for conductor wire 3 was abraded in the same location, consistent with the short circuit detected and reported signal issue.Further investigation concluded the black material found on electrode 3 to be a biological foreign material.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the short circuit is consistent with damage during use.
|