• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER 7.5X7.5CM INT; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER 7.5X7.5CM INT; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Catalog Number 66800269
Device Problem Material Disintegration (1177)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/29/2020
Event Type  malfunction  
Event Description
It was reported that on removal of the sacral dressing, some particles of silicone remained on the patient's skin surrounding the wound.On removal of the 15x15cm dressing, again, some particles of silicone remained on the patient's skin surrounding the wound.The wounds were both cavities and there was risk the silicone could end up in the wounds.These occurred on different patients, on different wards and different sites within the trust.On opening a new dressing 7.5x7.5cm, parts of the silicone remained on the film cover.It is unknown how many patients are involved.No patient harm reported.
 
Manufacturer Narrative
The device, used in treatment, has not been received for evaluation.No samples or further details were supplied, therefore we are unable to establish a relationship between the device and the event reported or determine a root cause on this occasion.No batch/lot details have been provided, due to this we are unable to conduct a review of the device history.However, at this time we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history for the reported event has been reviewed revealing further instances in the past three years.The wound contact surface is coated with a gentle silicone adhesive layer that ensures non-traumatic removal at dressing changes.The silicone adhesive will feel sticky when compared to an acrylic adhesive and is specially designed to be very soft and gentle, but can soften further, particularly at higher temperatures.This is an inherent characteristic of all of the silicone adhesives and gives them their soft / gentle properties.It is therefore possible if the product has been stored at a high temperature, this could have contributed to the reported issue.No further actions are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith + nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN GENTLE BORDER 7.5X7.5CM INT
Type of Device
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10376728
MDR Text Key201990256
Report Number8043484-2020-02319
Device Sequence Number1
Product Code NAD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800269
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-