ST. JUDE MEDICAL, INC. ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100014514
Device Problems Failure of Device to Self-Test (2937); Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/05/2020
Event Type  Injury  
Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During an ablation procedure, after ablation had been started the error message "amplifier error, log out and power cycle both the ensite velocity dws and amplifier before restarting.Contact sjm technical support" was displayed and the amplifier status light turned orange.The systems were power cycled and upon restarting ablation the same issue occurred.Troubleshooting was not able to resolve the issue and the procedure was cancelled.There were no adverse consequences to the patient.
 
Manufacturer Narrative
One ensite velocity¿ system velocity amplifier was received into the lab for analysis.Visual inspection of the returned product confirmed all input and output connectors are free of physical damage and all labeling is legible and correctly oriented.Power was applied to the returned amplifier which successfully completed the post (power on self-test) and the system status light changed to green.System log files were reviewed which indicate the device has not been used since refurbishment.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information provided to abbott and the investigation performed, the root cause of the reported event cannot be conclusively determined as no abnormalities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key10378249
MDR Text Key202004331
Report Number2184149-2020-00120
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number100014514
Device Catalogue Number100014514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Was the Report Sent to FDA? No
Date Manufacturer Received09/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-