AESCULAP AG PLASMACUP DC SIZE 50MM; HIP ENDOPROSTHETICS Back to Search Results
Model Number NH550T
Device Problem Use of Device Problem (1670)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2020
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.This is a similar device report.A similar device was sold to us.
 
Event Description
It was reported that there was an issue with a plasmacup.According to the complaint description the implants involved in this complaint could not be fixed during surgery.During the total hip replacement (thr) the reamer instrument (fs960r and fs974) was damaged and was not fixed and wandered.The acetabulum cup portion has been reamed more than previously planned.Although nh548t was implanted first, it was replaced with nt550t because it was not fixed.All implants were removed because they were probably not fixed.An other implant of a competitor company was used.A pubis fracture occured during surgery.An additional medical intervention was necessary.Additional information was not provided nor available.The adverse event is filed under aag references (b)(4).Associated medwatch-reports: 9610612-2020-00405 (400480311 - nh550t) similar device sold to us.9610612-2020-00406 (400480315 - nh548t).9610612-2020-00407 (400480317 - nh522d).
 
Manufacturer Narrative
Additional information: two involved components were added to this complaint: fs960r - orthopilot tha reamer shank mpr harris - lot unknown.Fs974 - gloves protection f/fs959r/fs960r/fs961r - lot unknown.Associated medwatch-reports: (b)(4), 9610612-2020-00406 (b)(4), 9610612-2020-00407(b)(4).
 
Manufacturer Narrative
Based upon new information received, this event was re-evaluated and is considered no longer reportable - no malfunction or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASMACUP DC SIZE 50MM
Type of Device
HIP ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10379944
MDR Text Key207959846
Report Number9610612-2020-00405
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNH550T
Device Catalogue NumberNH550T
Device Lot Number52521295
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FS960R LOT UNKNOWN.; FS974 LOT UNKNOWN.
Patient Outcome(s) Required Intervention;
-
-