• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG PLASMACUP DC SIZE 48MM; HIP ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG PLASMACUP DC SIZE 48MM; HIP ENDOPROSTHETICS Back to Search Results
Model Number NH548T
Device Problem Use of Device Problem (1670)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2020
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.This is a similar device report.A similar device was sold to us.
 
Event Description
It was reported that there was an issue with a plasmacup.According to the complaint description the implants involved in this complaint could not be fixed during surgery.During the total hip replacement (thr) the reamer instrument (fs960r and fs974) was damaged and was not fixed and wandered.The acetabulum cup portion has been reamed more than previously planned.Although nh548t was implanted first, it was replaced with nt550t because it was not fixed.All implants were removed because they were probably not fixed.An other implant of a competitor company was used.A pubis fracture occured during surgery.An additional medical intervention was necessary.Additional information was not provided nor available.The adverse event is filed under aag references (b)(4).Associated medwatch-reports: 9610612-2020-00405 (400480311 - nh550t) similar device sold to us.9610612-2020-00406 (400480315 - nh548t).9610612-2020-00407 (400480317 - nh522d).
 
Manufacturer Narrative
Additional information: two involved components were added to this complaint: fs960r - orthopilot tha reamer shank mpr harris - lot unknown.Fs974 - gloves protection f/fs959r/fs960r/fs961r - lot unknown.Associated medwatch-reports: 9610612-2020-00405 (400480311 - nh550t) similar device sold to us, 9610612-2020-00407 (400480317 - nh522d).
 
Manufacturer Narrative
Based upon new information received, this event was re-evaluated and is considered no longer reportable - no malfunction or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASMACUP DC SIZE 48MM
Type of Device
HIP ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10380227
MDR Text Key207960106
Report Number9610612-2020-00406
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNH548T
Device Catalogue NumberNH548T
Device Lot Number52543977
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/08/2020
Initial Date FDA Received08/07/2020
Supplement Dates Manufacturer Received07/08/2020
11/02/2020
Supplement Dates FDA Received08/12/2020
04/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FS960R LOT UNKNOWN.; FS974 LOT UNKNOWN.
Patient Outcome(s) Required Intervention;
-
-