Visual inspection: the rib on the top of the screw was damaged and deformed, most likely from being driven through the locking ring on the plate.Device and complaint history records were reviewed, and no relevant manufacturing issues or similar complaints were identified.From the surgical technique: the amount of torque required to complete final tightening can be done with a single hand, and should not exceed one quarter turn once the screw is underneath the ring.In addition to the tactile sensation of the locking ring closing over the bone screw head, final screw locking should also be confirmed visually with the ring being clearly visible over the bone screw head.It is unknown how many turns were applied after the locking ring became visible.The damage on the screw is consistent with being driven below the locking ring.Most likely root cause is excessive torque or too many turns applied to the screw once the locking ring became visible.
|