The initial reporter stated they received discrepant results for two patient samples tested with ise indirect k for gen.2 on a cobas 6000 c (501) module.The initial values were reported outside of the laboratory.The reporter stated there are two c 501 systems in the same line, serial numbers (b)(4) and (b)(4).The reporter could not confirm which serial number was used to test the affected samples.The reporter mentioned they received ise noise alarms on c 501 serial number (b)(4).A noise alarm was received when calibrating this analyzer on the morning of (b)(6) 2020.The calibration was repeated and there were no further alarms.Controls recovered fine.The first sample initially resulted with a k value of 8 mmol/l, which repeated as 3.9 mmol/l.The second sample initially resulted with a k value of 6 mmol/l, which repeated as 3 mmol/l.The k electrode lot number and expiration date were requested, but not provided.
|