• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES(IRELAND) CUSA EXCEL 36KHZ STRAIGHT HANDPIECE; ULTRASONIC SURGICAL PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES(IRELAND) CUSA EXCEL 36KHZ STRAIGHT HANDPIECE; ULTRASONIC SURGICAL PRODUCTS Back to Search Results
Catalog Number C2602
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/16/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not yet received by the manufacturer for evaluation.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A customer reported that no vibration was detected on the c2602 excel 36khz straight handpiece and alarms were triggered during vibration test on (b)(6) 2020.No patient injury was reported and no known delay in surgery.Additional information has been requested.
 
Manufacturer Narrative
Additional information: d10, g4, g7, h2, h3, h4, h6, h10.Device identifier: (b)(4).Product identifier: (b)(4).The device was returned for evaluation.The device history record (dhr) was reviewed and showed no anomalies related to the reported failure.The reported complaint was confirmed based on low amplitude, reported vibration failure, and transducer failure.This complaint is consistent with transducer delamniation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSA EXCEL 36KHZ STRAIGHT HANDPIECE
Type of Device
ULTRASONIC SURGICAL PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
MDR Report Key10389082
MDR Text Key205330876
Report Number3006697299-2020-00092
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
PMA/PMN Number
K141668
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC2602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Initial Date Manufacturer Received 07/15/2020
Initial Date FDA Received08/10/2020
Supplement Dates Manufacturer Received09/07/2020
Supplement Dates FDA Received09/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-