• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12, STERILE,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12, STERILE, Back to Search Results
Model Number WA22302D
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bowel Perforation (2668)
Event Type  Injury  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, patient's outcome the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transvaginal uterine fibroid resection procedure at an unknown date around 2018, it is suspected that the patient has suffered an intestinal injury although the muscle layer was scraped and no obvious perforation was made.The intended procedure was completed with an intervention procedure on the next day when the injury of the intestine was sutured.No further information was provided and there was no report of malfunction for any of the olympus medical devices.
 
Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation/investigation since it was reportedly discarded by the user facility.There were no reports of any malfunction of the hf-resection electrode.Based on the information available, the exact cause of the reported phenomenon and the patient¿s outcome could not be determined and is being judged as unknown.Furthermore, a dhr review could not be performed since basic data of article identification (lot number) are missing.Instead a manufacturing and quality control review was performed for the last 24 months of production.There were no non-conformities or deviations regarding the described issue.The case will be closed on olympus side with no further actions.However, the reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12, STERILE,
Type of Device
HF-RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key10390455
MDR Text Key210021290
Report Number9610773-2020-00185
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051665
UDI-Public14042761051665
Combination Product (y/n)N
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWA22302D
Device Catalogue NumberWA22302D
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/04/2020
Initial Date FDA Received08/11/2020
Supplement Dates Manufacturer Received10/20/2020
Supplement Dates FDA Received10/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-