The reported events of inappropriate shock and noise oversensing were confirmed in the laboratory.The stored electrogram was reviewed and noise was observed on the ventricular sensing channel.However, there were no viewable stored electrograms showing delivered therapies.The set screws were tested, and all set screws were able to be tightened down and secure test leads normally.It was reported that the right ventricular lead came out of the device header while removing the device from the patient.The cause of the reported event appeared consistent with poor set screw lead connection in the header that was procedure related.The device was tested on the bench, and no anomalies were found.
|