• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2020
Event Type  malfunction  
Event Description
The recipient reportedly experienced sound quality issues.The recipient was a non-user of the device and elected revision surgery.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed a damaged antenna coil.This is not believed to be related to the return reason.System lock was verified.The device passed the electrical and mechanical tests performed.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Additional information: section d.10.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10398620
MDR Text Key203473981
Report Number3006556115-2020-00792
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 07/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/31/2013
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/12/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received09/04/2020
10/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age13 YR
-
-