Model Number CI-1400-01 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/10/2020 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced sound quality issues.The recipient was a non-user of the device and elected revision surgery.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed a damaged antenna coil.This is not believed to be related to the return reason.System lock was verified.The device passed the electrical and mechanical tests performed.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Additional information: section d.10.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|