• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE Back to Search Results
Catalog Number 130760000
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Event Date 07/15/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during a shoulder prothesis revision there was a delay of surgery with a delta xtend prothesis.The implant was broken and the surgeon had removed the device.The implantation date: 2016.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information stated that the patient experienced pain for about 18 months, functional impotence and lameness.X-ray result revealed disassembly of the glenoid sphere.There was a surgical delay of 60 minutes.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned devices confirmed the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.The glenosphere implant components were found to be disassociated from themselves.The screw of the glenosphere was broken off inside the metaglene.There was evidence of disassociation of the glenosphere from the metaglene.The poly cup was found to be fractured and worn.There were no visible defects with the locking and non-locking screws.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4) investigation summary
=
> examination of the returned devices confirmed the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.H10 additional narrative: added: h6 (re-captured codes) corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND METAGLENE
Type of Device
DELTA XTEND IMPLANTS : SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex IN 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex IN 69801
FR   69801
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr
warsaw, IN 46581-0988
6107428552
MDR Report Key10401344
MDR Text Key202716275
Report Number1818910-2020-18030
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K192448
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130760000
Device Lot Number5257263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/30/2020
Initial Date FDA Received08/13/2020
Supplement Dates Manufacturer Received08/28/2020
09/15/2020
09/30/2020
03/20/2023
Supplement Dates FDA Received08/31/2020
09/22/2020
10/06/2020
03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND GLENOSPHERE STD D42MM; DXTEND MODULAR HUM STEM D14 HA; DXTEND SCREW LOCK D4.5X36MM; DXTEND SCREW NO LOCK D4.5X24MM; DXTEND STAND PE CUP D42 +3MM; UNKNOWN SHOULDER LOCKING SCREW; DXTEND GLENOSPHERE STD D42MM; DXTEND MODULAR HUM STEM D14 HA; DXTEND SCREW LOCK D4.5X36MM; DXTEND SCREW NO LOCK D4.5X24MM; DXTEND STAND PE CUP D42 +3MM
Patient Outcome(s) Required Intervention;
-
-