Device Problem
Power Problem (3010)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
It was reported that before surgery that an electric dermatome would intermittently stop.There was no harm and no delay reported.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor was malfunctioning and the motor seal and insulator were damaged.The motor, seal, and insulator were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|
Event Description
|
No additional event information available.
|
|