Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.A field service engineering (fse) was at the customer's site to address reported event.Fse visually confirmed reported error on computer screen which states "the end marker was encountered before the end of the sample rack, possible wrong position".Fse verified proper operation and alignment of the end maker sensor, reinstalled software and version up, then verified the error was no longer appearing during normal operation.Fse successful ran quality controls without error, and results were within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A complaint and service history review for similar complaints was performed for aia-2000 serial number (b)(4) from installation date through aware date.There were no other similar complaints identified during the searched period.The most probable cause of the reported event was due to the software needed reloading.
|
A customer reported getting a computer error during quality control run on the aia-2000 instrument.The customer attempted rebooting the computer and instrument, but error persisted.A field service engineer was dispatched to address the reported event, which resulted in delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|