• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number UNK-P-IPP
Device Problems Failure to Cycle (1142); Mechanical Problem (1384); Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/02/2020
Event Type  Injury  
Event Description
It was reported that the patient experienced mechanical issues with an inflatable penile prosthesis (ipp) pump as the fluid would not pump.A replacement surgery was performed in which it was noticed that the tubing going to the right cylinder was broken at the base of the pump.The patient did not experience any adverse events and had an excellent outcome following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
alyson harris
10700 bren road w
minnetonka, MN 55343
4083953452
MDR Report Key10407689
MDR Text Key202930343
Report Number2183959-2020-03518
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNK-P-IPP
Device Catalogue NumberUNK-P-IPP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/24/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age71 YR
-
-