• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (wires exposed) has been confirmed.Upon investigation the cable connecting the distribution node (dn) to the rear therapy electrodes was pulled from the strain relief, damaging wires in the cable.The root cause for the strained cable was excessive force.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor reported that a patient's electrode belt had exposed wires.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, pa 
9683333
MDR Report Key10409403
MDR Text Key203490456
Report Number3008642652-2020-06438
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Date Manufacturer Received07/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-