• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAKO NORTH AMERICA, INC. AUTOSTAINER LINK 48; AUTOSTAINER LINK 48 INSTRUMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAKO NORTH AMERICA, INC. AUTOSTAINER LINK 48; AUTOSTAINER LINK 48 INSTRUMENT SYSTEM Back to Search Results
Model Number AS480
Device Problems Fluid/Blood Leak (1250); Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2020
Event Type  malfunction  
Event Description
As part of agilent's continuous improvement process, our procedures were reviewed and it was determined that there is a need to report potential risks associated with critical liquid dispense system components on the autostainer.We have assessed all autostainer dispense and leakage complaints from november of 2017 through march of 2020 that resulted in alteration in staining or potential alteration in slide staining.Following this review, it was decided to retrospectively file an mdr for customer complaint (b)(4).This is an initial report.Summary: based on complaint report or investigated failure mode, there was potential for a staining alteration.Customer complaint record reported the event as follows: did not aspirate well, air in the tube no direct or indirect patient harm or user harm have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOSTAINER LINK 48 INSTRUMENT SYSTEM
Manufacturer (Section D)
DAKO NORTH AMERICA, INC.
6392 via real
carpinteria, ca
Manufacturer (Section G)
DAKO NORTH AMERICA, INC.
6392 via real
carpinteria, ca
Manufacturer Contact
mani siravanta
6392 via real
carpinteria, ca 
5665462
MDR Report Key10411037
MDR Text Key203651094
Report Number2022180-2020-00208
Device Sequence Number1
Product Code KPA
UDI-Device Identifier05700572035497
UDI-Public05700572035497
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/28/2020
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-