According to the reporter, during esophagectomy/gastric tube, after using the clip device several times, the clip was not loaded and the counter was not displayed.The surgery was finished using a new product.There was no patient injury.
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The reported cond ition of the clips not loading properly was not confirmed.The most likely cause could not be identified, no related problem was detected with the device.The instrument handle was disassembled for visualization of internal components, revealing a proper handle as sembly.Visual inspection of the clip counter noted particulate matter inside the contacts preventing activation of the counter.The instrument was applied to test media and all clips were fired.Functionally, the clips loaded into the jaws, formed properly, and r emained securely attached to test media.It was reported that the digital reader did not work properly.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|