MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFD0433
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Embolus (1830); Perforation of Esophagus (2399)
Event Date 08/06/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.In addition, no log files were received.Review of the device history record was not possible as the lot number is unknown.Based on the information received, the cause of the reported embolism and atrio-esophageal fistula could not be conclusively determined.
 
Event Description
Five weeks post pulmonary vein isolation procedure, a cerebral embolism and atrio-esophageal fistula were noted.A pulmonary vein isolation procedure was performed on (b)(6) 2020 with no complications or patient symptoms.Two or three weeks following the procedure, a stent was placed.Two weeks later, the patient presented to the hospital disoriented and was sent for a ct which revealed an air embolism in the sulci.Subsequently, the ct of the chest revealed the fistula.The patient was taken to the hyperbaric chamber for two sessions of treatment.The patient is in icu in unstable condition.There were no performance issues with any abbott devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key10414662
MDR Text Key203157903
Report Number3005334138-2020-00359
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial
Report Date 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAFD0433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient Weight91
-
-