Model Number CI-1601-05 |
Device Problems
Mechanical Problem (1384); Output Problem (3005)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/29/2020 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing decreased performance.Programming adjustments were made, however, the issue did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical tests performed.The scanning electron microscopy (sem) analysis revealed corrosion on some electrode wires.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Search Alerts/Recalls
|