Model Number 38270
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Patient Involvement (2645)
Event Date 01/16/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by manufacturer: returned product consisted of an imager diagnostic catheter in the un-opened sterile pouch.Tensile strength inspection of the returned product was performed and the device failed to meet specification for tip detachment.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 23jul2020.A 5f bern imager ii angiographic catheter was returned to bsc from a customer with no reported complaint against the device.Device analysis on the returned device found that the device failed tensile testing leading to a tip detachment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMAGER II ANGIOGRAPHIC CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
TFX ENGINEERING LIMITED T/A
unit 7 8 and 9
annacotty li
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10417812
MDR Text Key203300208
Report Number2134265-2020-10972
Device Sequence Number1
Product Code DQO
UDI-Device Identifier08714729354871
UDI-Public08714729354871
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K121694
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/12/2020
Device Model Number38270
Device Catalogue Number38270
Device Lot Number0000134600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/23/2020
Initial Date FDA Received08/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number92484513-FA
Patient Sequence Number1
-
-