• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3369-40Q
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Patient Involvement (2645)
Event Date 08/02/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that prior to implant procedure a new implantable cardioverter defibrillator was removed from the package and found to be in backup operation.There was no patient involvement.
 
Manufacturer Narrative
The reported event of reset was confirmed in the lab.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.The device passed memory testing as expected.Bench testing was performed, which includes impedance, sensing, pacing, and high voltage output tests and the device was normal.A parity error was seen to be the cause of the backup vvi.The reset could not be reproduced in the lab; hence the root cause of the error remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10417952
MDR Text Key203279276
Report Number2017865-2020-11692
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000093434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-