Model Number CD3369-40Q
Device Problem Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/30/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2017865-2020-11461.It was reported that the patient presented in clinic for ablation procedure.Interrogation of the patient's implantable cardioverter defibrillator (icd) revealed that the patient's right ventricular lead exhibited under-sensing during a supraventricular tachycardia episode.It was also reported that radiofrequency (rf) telemetry was initially established but was lost and the icd only allowed for wand or inductive telemetry.No intervention was performed.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10418748
MDR Text Key203278730
Report Number2017865-2020-11463
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000088980
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE
Patient Age59 YR
-
-