The sales representative reported on behalf of the customer that the pezs01 was being used during a hip arthroscopy, labral repair, femoral neck osteoplasty on (b)(6) 2020 when the seal dislodged and was pushed into the joint.The user was able to retrieve the component using a grasper.This caused a 1-minute delay in the procedure.The procedure was completed as planned using the same device as the procedure was at completion when the event occurred.There was no report of harm to the user or the patient.There was no medical intervention or hospitalization for the patient due to the event.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
The device will not be returned for evaluation however the provided photographic evidence exhibits the reported event.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 3 complaints, regarding 3 devices, for this device family and failure mode.During this same time frame 961 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.003.Per the instructions for use, the user is advised the following: warnings: preoperative and operating procedure, including knowledge of surgical techniques and proper selection and placement of implants are important considerations in the successful utilization of these devices.Precautions: inspect instruments prior to use to ensure they are in good physical condition and function properly.There should be no loose, broken or misaligned parts.Exercise care in the user of these devices to minimize side or bending loads.Do not use excessive force on instruments to avoid damage or breakage during insertion and use.Use care while passing sharp instrumentation to prevent damage to the device.Inspect instruments after use to ensure they have not been damaged.It is the responsibility of the surgeon to become familiar with the proper techniques for use.Maintain proper alignment during insertion.This issue will continue to be monitored through the complaint system to assure patient safety.
|