• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML Back to Search Results
Model Number CSE-P-125-JA
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/25/2020
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2020 manufacturing performed a visual inspection of the received 125 ml bowl from the cell saver set and confirmed three (3) cracks at union between base and inner bowl core which resulted in the error the customer received.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
 
Event Description
On (b)(6) 2020 haemontics (b)(6) was informed by the customer (b)(6) for children and adults during cardiac surgery, an error of "poor blood return (long blood return)" was displayed and occured in the blood return of the first cycle.When the centrifuge bowl was checked, an inner core leak had occurred while utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 125ml.There was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer Contact
miranda holland
125 summer street
boston, ma 
8487100
MDR Report Key10421277
MDR Text Key209225952
Report Number1219343-2020-00075
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/27/2022
Device Model NumberCSE-P-125-JA
Device Lot Number0819082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/06/2020
Initial Date FDA Received08/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-