Model Number CSE-P-125-JA |
Device Problems
Crack (1135); Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
On (b)(6) 2020 haemonetics manufacturing performed a visual inspection of the received 125 ml bowl from the cell saver elite set and confirmed blood in the inner core with 1 bowl crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing evaluation due the event of the past report.
|
|
Event Description
|
On (b)(6) 2020 haemonetics (b)(6) was informed by customer (b)(6) medical center while utilizing the cell saver® elite® plus system and cell saver® elite set - 125ml during the bood return phase at the end of the bentall total-arch openstent procedure they received an error "long blood return".When the bowl was removed, blood was flowing inside the core.The bowl was replaced with a new one.The onset cycle was about 15 to 16 during and after surgery.There was no reported impact to patients' health.
|
|