Model Number CSE-P-125-JA
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/03/2020
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2020 haemonetics manufacturing performed a visual inspection of the received 125 ml bowl from the cell saver elite set and confirmed blood in the inner core with 1 bowl crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing evaluation due the event of the past report.
 
Event Description
On (b)(6) 2020 haemonetics (b)(6) was informed by customer (b)(6) medical center while utilizing the cell saver® elite® plus system and cell saver® elite set - 125ml during the bood return phase at the end of the bentall total-arch openstent procedure they received an error "long blood return".When the bowl was removed, blood was flowing inside the core.The bowl was replaced with a new one.The onset cycle was about 15 to 16 during and after surgery.There was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE PLUS AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer Contact
miranda holland
125 summer street
boston, ma 
8487100
MDR Report Key10421320
MDR Text Key205685284
Report Number1219343-2020-00080
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/13/2023
Device Model NumberCSE-P-125-JA
Device Lot Number0120047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2020
Initial Date FDA Received08/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-