Model Number 1458Q/86 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This product is registered as a combination product.The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported that the patient presented for an implant procedure.During the procedure, the left ventricular lead could not be fixed.The lead was removed and replaced.The patient was in stable condition.
|
|
Manufacturer Narrative
|
Analysis found that a complete lead was returned in one piece measuring 86.4cm.Analysis was normal.No anomalies were found.
|
|