Model Number CD3365-40Q
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 07/30/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient experienced an inappropriate shock from the implantable cardioverter defibrillator and presented to the emergency room for a follow up.Upon examination, it was determined that the device was programmed inadequately where the atrial rhythm fell into a blanking period and was not sensed by the device.Subsequently, the ventricular rhythm was sensed at a higher rate causing the inappropriate high voltage therapy.The device was reprogrammed to resolve the anomaly.The patient was reported to be in normal condition.No further incident was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10423904
MDR Text Key203443759
Report Number2017865-2020-11786
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508292
UDI-Public05414734508292
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2017
Device Model NumberCD3365-40Q
Device Catalogue NumberCD3365-40Q
Device Lot NumberA000015171
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age65 YR
-
-