On (b)(6) 2019, a procedure was performed to implant one (1) stimq peripheral nerve stimulator (pns) system.Receiver (fr8a-rcv-a0) at the occipital nerve for head pain.The stimq pns system is not indicated for use at occipital.Nerves and was implanted off-label at the clinician's discretion.The territory manager confirmed that the implant.Procedure was performed in a sterile environment, sterile field handling protocols were used, the procedure was completed in accordance with the product instructions for use, and the sterile barriers of all product used were intact prior to implant.On (b)(6) 2019, the patient visited the implanting clinician's office for follow up and the implanting clinician observed that the stimulator was protruding through the skin.At the time of the event, the patient was still receiving therapy.The implanting clinician made the decision to explant the device as a precaution.The device was explanted the same day and no further complications were reported.Based on this information, the device erosion was confirmed/replicated, there is no evidence that product did not meet specification as this was used off label and the stimulator is used for treatment of pain.The cause of the device erosion is due to "off label use"-user error.Device history record was reviewed and there were no non-conformances and product met specification at final release.
|