Model Number TR-H200L-S
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Type  Injury  
Manufacturer Narrative
A visual examination of the returned radial head was performed under magnification.The radial head has a clean, smooth surface inside and out, without any wear or discoloration.Additional mdr associated with this event: 3025141-2020-00196; stem.
 
Event Description
An arh radial head replacement system was implanted in the patient two years ago.Post op, the patient developed an allergic reaction that moved throughout the body so the hardware was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20.0MM HEAD, LEFT
Type of Device
PROSTHESIS, ELBOW, HEMI-, RADIAL: HEAD
Manufacturer (Section D)
ACUMED LLC
5885 ne cornelius pass road
hillsboro, or
Manufacturer (Section G)
ACUMED LLC
5885 ne cornelius pass road
hillsboro, or
Manufacturer Contact
micki anderson
5885 ne cornelius pass road
hillsboro, or 
6279957138
MDR Report Key10425668
MDR Text Key203502911
Report Number3025141-2020-00195
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131845
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTR-H200L-S
Device Catalogue NumberTR-H200L-S
Device Lot Number421449
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2020
Date Manufacturer Received07/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-