Model Number M3-70A |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Device has been returned for evaluation.The customer¿s complaint of cloudy image was confirmed.It was observed it was due loose lens and debris.The meniscus was out of place causing debris.The root cause cannot be determined.Olympus will continue to monitor the field performance of this device.Dhr review was performed and no non-conformances that would cause the reported malfunction were noted.No further information was reported.No further information was reported.
|
|
Event Description
|
The customer reported to olympus that the device's image was cloudy.There was no patient injury reported.
|
|
Manufacturer Narrative
|
Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
|
|