• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC M3-GOLD AUTOCLAVABLE LATERAL TELESCOPE 70 DEG; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC M3-GOLD AUTOCLAVABLE LATERAL TELESCOPE 70 DEG; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number M3-70A
Device Problem Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device has been returned for evaluation.The customer¿s complaint of cloudy image was confirmed.It was observed it was due loose lens and debris.The meniscus was out of place causing debris.The root cause cannot be determined.Olympus will continue to monitor the field performance of this device.Dhr review was performed and no non-conformances that would cause the reported malfunction were noted.No further information was reported.No further information was reported.
 
Event Description
The customer reported to olympus that the device's image was cloudy.There was no patient injury reported.
 
Manufacturer Narrative
Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M3-GOLD AUTOCLAVABLE LATERAL TELESCOPE 70 DEG
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key10429019
MDR Text Key208336045
Report Number1519132-2020-00054
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
PMA/PMN Number
K890328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3-70A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/30/2020
Initial Date FDA Received08/19/2020
Supplement Dates Manufacturer Received03/21/2021
Supplement Dates FDA Received03/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-