Catalog Number 66800039
Device Problem Connection Problem (2900)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2020
Event Type  malfunction  
Event Description
It was reported that the hand-piece interface is not in the correct position.When the hand-piece is attempted to be inserted, it won't turn to the active/on position, as the piston appears to be stuck in the 'extended' position.The sales rep made several attempts to clear the piston, by deploying the foot pedal in quick bursts, but to no avail.The machine likely requires a quick look and repair from engineering.No patient was involved.
 
Manufacturer Narrative
The device that was intended for use in treatment was returned for evaluation.There was a relationship between the reported event and the device.A visual inspection was performed on the exterior of product and no physical damage was observed.Functional inspection was performed and showed the test pump cartridge was difficult to turn to the locked position due to corrosion inside u.I.Assembly.The root cause is determined to be caused by corrosion.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.Smith and nephew will continue to monitor for any adverse trends relating to this product.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10431303
MDR Text Key203672373
Report Number8043484-2020-02551
Device Sequence Number1
Product Code FQH
UDI-Device Identifier04582111153524
UDI-Public4582111153524
Combination Product (y/n)N
PMA/PMN Number
K110958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,user facility
Type of Report Initial,Followup
Report Date 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number66800039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Date Manufacturer Received04/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-