The reported event of inappropriate therapy due to oversensing was not replicated in the laboratory.The device¿s sensing, pacing, lead impedance, high voltage charging, high voltage delivery and high voltage shock impedance were tested, and the device¿s function was normal.The device¿s associated lead was returned and analyzed, and no anomaly was found.The device¿s stored egms were reviewed and an egm recorded on (b)(6) 2020 showed p-waves being sensed on the ventricular sensing channel; this is consistent with lead dislodgement which was reported in 2938836-2020-07883.
|